国产精品久久久久久久免费-国产精品久久久久久久人热-国产精品久久久久久久网站-国产精品久久久久久久专区-国产精品久久久久久免费

堡盟電導率變送器專為介質分離與分析而設計,適用于食品飲料及水處理行業的諸多應用。它在精度和圖形化顯示方面有著非常突出的優勢。


Operating principle of an inductive conductivity sensor Baumer AFI

Differentiation of conductive and inductive technology

The classic set up of conductive conductivity sensors, whose electrodes are in galvanic contact to the medium to be measured, is limited by the point at which high ion concentrations may result in the so-called polarization effect. This can constitute an additional resistance, distorting the measurement result. Similarly, deposits (e.g. sodium hydroxide) can create insulating layers that make conductive conductivity measurement impossible.

For applications with high ion concentrations, such as leaches and acids with conductivity values in the range of 100 mS/cm and when there is a risk of deposits, inductive technology offers the only reliable conductivity measurement and thus reliable control of processes. However, the inductive measurement principle is not suited for measuring very small conductivity values. The smallest measurement range is 500 µS/cm (0.5 mS/cm); in this range precise measurements of around 50 µS/cm can still be carried out.

Design of the inductive conductivity sensor AFI
Teaser_Service_Support_CIP_Zchn-7040_600x338.jpg

The sensor element with a seamless PEEK housing contains two ring core coils that act as two virtual transformers arranged in series. The first primary winding is fed by an oscillator in the kilohertz range. The liquid loop that is created by the current measuring passage through the interior of the two ring core coils and the surrounding area connects the secondary side of the first transformer to the primary side of the second transformer. This connection can be considered a common winding of both transformers. The secondary winding of the second transformer is connected to a galvanometer. The volume of the liquid resistance RM results in an associated current measurement value. Using a simple calculation rule (Ohm’s law), this value together with the known oscillator voltage UG is transformed into the conductance value GM to be output.

The sensitivity of the galvanometer must be adjusted for different conductance value ranges. The user can do this by setting freely defined measurement ranges that can be selected during operation by activating control inputs.

The inductive measurement principle is not reliant on the movement of liquids in the current measuring passage. Nevertheless, it is recommend to align the channel in the direction of the flow to achieve better cleaning results. The symmetrical design of the channel accommodates both flow directions without the risk of clogging by media components.

A precise and highly responsive Pt100 temperature sensor is integrated into the tip of the sensor element. The temperature of the medium measured by this is available as a measuring signal and is used in addition to the physical temperature compensation of the raw conductivity value.
 


溫度補償

液體的電導率一般與溫度密切相關。溫度每升高1K,許多水溶液的電導率會增加2%。為了測量結果的可比性,測量設備會將直接測得的電導率換算成參考溫度下對應的電導率(一般為25 °C)。此外,AFI變送器還允許溫度系數以%/K為單位直接輸入。如無需進行溫度補償,溫度系數必須定為0%/K。非線性溫度系數可額外使用一個二次項來加以定義。

Teaser_Know-how_E-Zchn-6576_600x338.jpg
回到頂部
主站蜘蛛池模板: 免费一级a毛片在线播| 久久免费观看国产精品| 青青国产成人久久激情911| 成人小视频在线| 手机看片亚洲| 久久国产这里只有精品| 色屁屁一区二区三区视频国产| 美国毛片毛片全部免费| 国产普通话自拍| 亚洲精品乱码国产精品乱码| 日本tv欧美tv天堂| 免费国产成高清人在线视频| 视频在线观看一区二区三区| 国产观看精品一区二区三区| 日本亚洲黄色| 2021国产精品一区二区在线| 日韩特级黄色片| 伊人久热这里只有精品视频99| 国产私拍写真福利视频| 激情区| 久久青青国产| 欧美日韩综合网在线观看| 2015日韩永久免费视频播放| 三级在线国产| 中文字幕在线国产| 色婷婷在线视频| 免费看的毛片| 国产成人免费全部网站| 黄色网址在线看| 国产精品大尺度尺度视频| 欧美日韩你懂的| 青青草在线视频免费观看| 一个色综合网| jizz免费软件| 日本xxxwww色视频| 亚洲毛片儿| 色婷婷综合久久久| 色婷婷综合久久久| 成人黄网大全在线观看| 欧美色频| 国产大毛片|